# Procedure B2: Connect a plate reader via Canopy :::{admonition} Note A: Prerequisites :class: note This procedure assumes the following: - You have Labrador open and are logged in. If this is not the case, follow [Procedure A2, "Log in"](../account/login.md). - Your account is a member of an organization with an active license. To check this, follow [Procedure A4, "Check license status"](../account/check-license.md). ::: ## Connect the Canopy 1. Get your Canopy and the USB-C to USB-A cable provided with it. 2. Plug the USB-A end of the cable into the computer running Labrador. - Note: Ideally, this should be plugged _directly_ into your computer, rather than through a USB hub. While most USB hubs work, some do not provide adequate power, and others can cause unpredictable behavior. If a hub must be used, it is the user's responsibility to ensure that no problems with device communication are apparent. (It is easiest to verify this using a directly-connected plate reader, rather than the Canopy.) 3. Plug the USB-C end of the cable into the Canopy. 4. Wait for the Canopy to finish booting, as indicated by the status LED no longer shining white or pulsing purple. (When the Canopy is first plugged in, it may take a few seconds for the status light(s) to turn on at all.) - If the Canopy status light turns red, try unplugging it from the USB cable and returning to step 3. If this continues to happen, contact [support@cerillo.bio](mailto:support@cerillo.bio). - If no lights turn on on the Canopy, and you are using a USB hub, it is likely that the hub is not providing adequate power to the Canopy. Try plugging it directly into your computer, or use a different USB hub. ## Power Up the Plate Reader 5. Get your plate reader and the USB-C to USB-A cable provided with it. If using AC mains power to run the plate reader: 6. Plug the provided USB-A AC adapter into a mains power outlet, and plug the USB-A end of the provided cable into the adapter. If using battery power to run the plate reader: 7. Plug the USB-A end of the provided cable into a Cerillo-approved battery pack (not provided with the plate reader). 8. Plug the USB-C end of the provided cable into the plate reader. 9. Wait for the plate reader to finish booting, as indicated by the status light no longer shining white. ## Load the Devices in Labrador 10. In Labrador, navigate to the **Devices** or **Home** screen using the link on the left-hand navigation menu (which may just appear as a **Home** icon if the Labrador window is small enough to collapse the sidebar). 11. Click the **Load devices** button. (If devices are already present, this button is at the top-right of the window.) 12. Wait for the spinner to disappear. This should take less than two minutes. 13. Confirm that both your Canopy and your plate reader are present in the resulting **Devices** screen. - If only your Canopy is present, try clicking **Load devices** again. Depending on environmental conditions, it may take two scans for the Canopy to find the plate reader. ```{eval-rst} .. raw:: latex \newpage ```