Experiment Setup
Main Experiment Settings
To configure and start a new experiment, click New experiment in the Device Home Page.

Number |
Element |
Description |
|---|---|---|
1 |
Experiment Name Input |
Name the experiment. Cannot use commas |
2 |
Wavelength select |
Select which wavelengths to measure. Only visible for multi-wavelength readers. Single wavelength readers only read in their specific wavelength. |
3 |
Experiment Type Selector |
Choose between Endpoint and Kinetic experiments. Click Endpoint to toggle to an endpoint experiment setup which will trigger the Endpoint Wizard when start is clicked. Endpoint readings must be enabled on the device through Device Settings for endpoint experiments to be enabled. Click Kinetic to enable kinetic experiments which will take measurements at a defined interval continuously or for a defined duration. When Kinetic is selected, the Interval, Continuous Read Toggle and Duration options will be visible. |
4 |
Interval Numerical Input |
User input that sets the interval between measurements in minutes. |
5 |
Duration Numerical Input |
User input to set the duration of the experiment. Toggle the Continue reading until stopped checkbox on to make the device read continuously until stopped by the user. Duration input will not apply if this is selected |
6 |
Plate Type Select |
Choose what type of plate is to be read. Note that selecting 6, 12, or Cerillo Co-Culture plate types will affect which detectors the plate reader uses to read wells. 96 well type will read all detectors and sensors. See plate mapping for more info on which sensors are used. |
7 |
Plate Layout |
Opens dialog to edit plate layout for the experiment. |
8 |
Use Template |
Choose a template to use for the plate layout. Changes can be made to the plate layout after a template has been copied. Changes made here will not affect the original template. |
9 |
Reset Defaults |
Certain inputs are cached between experiment runs to make it easier to run setup subsequent experiments. For instance, the plate type will default to the type used on most recent experiment run on the device. To reset these defaults to their original settings, you can click this button. |
10 |
Cancel Button |
Cancel experiment setup and return to the Device Home Page |
11 |
Start Button |
Begin experiment with values input in elements 1 - 5. If experiment is of type Endpoint the Endpoint Experiment Wizard will open to guide user through calibration, if necessary. If experiment is of type Kinetic and Kinetic auto-calibration is enabled in Device Settings, a prompt will ask you to place the plate in the reader before calibration starts automatically. Automated devices will show an Open device lid or Close device lid button in this prompt, and will show the same Lid is open dialog as the Endpoint Experiment Wizard if the lid is open when you click Ok. |
Multi-Wavelength Device Setup
For multi-wavelength devices (such as the Rayo Multi), a Wavelengths section appears in the experiment setup form. Each wavelength supported by the connected device is shown with a checkbox and its value in nanometers.
At least one wavelength must be checked before the experiment can be started. Any combination of the available wavelengths may be selected. The device will measure all checked wavelengths on every read cycle, and the results for each wavelength can be viewed independently in the Experiment Page.
Tip
Selecting more wavelengths increases the data collected per measurement but does not affect the read interval or duration settings.
Note
Stratus device’s 0-0.5 optical density range is not sensitive enough for most endpoint assays. The Stratus is best suited for kinetic experiments or endpoint experiments where the optical density is above 0.5.